ELECTRON MICROSCOPY
| Sample |
|---|
| BRISC complex bound to SHMT2 alpha |
| Sample Components |
| BRISC complex |
| SHMT2 alpha |
| Specimen Preparation | |
|---|---|
| Sample Aggregation State | PARTICLE |
| Vitrification Instrument | LEICA EM GP |
| Cryogen Name | ETHANE |
| Sample Vitrification Details | A 4 ul sample was applied to the grid and a protocol consisting of 30 s pre-blot incubation, 2 s blotting and no post-blot incubation was utilized for ... |
| 3D Reconstruction | |
|---|---|
| Reconstruction Method | SINGLE PARTICLE |
| Number of Particles | 35595 |
| Reported Resolution (Å) | 3.9 |
| Resolution Method | FSC 0.143 CUT-OFF |
| Other Details | |
| Refinement Type | |
| Symmetry Type | POINT |
| Point Symmetry | C2 |
| Map-Model Fitting and Refinement | |||||
|---|---|---|---|---|---|
| Id | 1 (6GVW, 5V7I) | ||||
| Refinement Space | REAL | ||||
| Refinement Protocol | FLEXIBLE FIT | ||||
| Refinement Target | Correlation coefficient and | ||||
| Overall B Value | 117 | ||||
| Fitting Procedure | |||||
| Details | Initial fitting with done using COOT, Rosetta was used for amino acid sequence threading and flexible fitting. COOT and ISOLDE were used for local reb ... | ||||
| Data Acquisition | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
| Electron Dose (electrons/Å**2) | 45 | ||||||||
| Imaging Experiment | 1 |
|---|---|
| Date of Experiment | |
| Temperature (Kelvin) | |
| Microscope Model | FEI TITAN KRIOS |
| Minimum Defocus (nm) | |
| Maximum Defocus (nm) | |
| Minimum Tilt Angle (degrees) | |
| Maximum Tilt Angle (degrees) | |
| Nominal CS | |
| Imaging Mode | BRIGHT FIELD |
| Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
| Nominal Magnification | 130000 |
| Calibrated Magnification | 58140 |
| Source | FIELD EMISSION GUN |
| Acceleration Voltage (kV) | 300 |
| Imaging Details |
| EM Software | ||
|---|---|---|
| Task | Software Package | Version |
| PARTICLE SELECTION | Gautomatch | 0.53 |
| IMAGE ACQUISITION | EPU | 1.1.0.30 |
| CTF CORRECTION | RELION | 2.1 |
| MODEL FITTING | Coot | 0.8.9 |
| MODEL FITTING | Rosetta | 2018.09.60072 |
| INITIAL EULER ASSIGNMENT | RELION | 2.1 |
| FINAL EULER ASSIGNMENT | RELION | 2.1 |
| CLASSIFICATION | RELION | 2.1 |
| RECONSTRUCTION | RELION | 2.1 |
| MODEL REFINEMENT | Rosetta | 2018.09.60072 |
| MODEL REFINEMENT | PHENIX | 1.15-3459 |
| Image Processing | ||||
|---|---|---|---|---|
| CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
| PHASE FLIPPING AND AMPLITUDE CORRECTION | CTF was determined using GCTF. CTF was corrected within RELION. | 332598 | Particles were auto-picked. | |